Skip to content

Add support for multiple olfactometers#24

Open
bruno-f-cruz wants to merge 15 commits into
mainfrom
feat-update-to-multiple-olfactometers
Open

Add support for multiple olfactometers#24
bruno-f-cruz wants to merge 15 commits into
mainfrom
feat-update-to-multiple-olfactometers

Conversation

@bruno-f-cruz
Copy link
Copy Markdown
Member

No description provided.

@bruno-f-cruz
Copy link
Copy Markdown
Member Author

@DaniAlcAlv I am going to leave this as is for now until you run further tests. Once you are happy with the results lmk so i can look at one of the datasets.
Thanks!

@bruno-f-cruz bruno-f-cruz added the blocked The issue is blocked by an external dependency label Apr 28, 2026
@DaniAlcAlv
Copy link
Copy Markdown

Here's the findings from the last calibration:

  • Olfactometer A Valve 2 is opening twice on both phases.
  • Olfactometer A Valve 1 Phase 2 only opens for a fraction of a second
  • The Olfactometer B Valve 3 Phase 2 test is not valid because the end valve closes too early at the end of the calibration.
  • During the second phase of the calibration, the baseline is too high. I’ll investigate why this is happening (might be due to backflow?).

Graph:
olf-calibration-2

Data: calibration_2026-05-19T015115Z.zip

@bruno-f-cruz
Copy link
Copy Markdown
Member Author

bruno-f-cruz commented May 20, 2026

Can you try with this configuration and report back?
olfactometer_rig.json

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment

Labels

blocked The issue is blocked by an external dependency

Projects

None yet

Development

Successfully merging this pull request may close these issues.