Skip to content

Latest commit

 

History

History
13 lines (10 loc) · 1.01 KB

File metadata and controls

13 lines (10 loc) · 1.01 KB

MedicalO2Arduino

Arduino codes and docs for using O2 medical sensors

Datasheet

KE-25_KE-50_DataSheet.pdf

Schematic

Circuit

Graph between KE-50 oxygen sensor and O2 multimeter

On this graph, we can see the curves of the O2 sensor KE-50 and the O2 multimeter CY-12C before and after they receive a O2 pulse of 1 second. After the pulse, we see the curve of the O2 multimeter decrease while the one for the O2 sensor KE-50 do not decrease yet because of an air exchange smaller than 200-300ml per minute. Sensors should be used under conditions where the air exchange is greater than 200-300ml per minute in order to obtain the response speed as specified in the response time graph on the datasheet. Graph