HP34401A config method sets most other values to default (address in documentation) #23

@WJohnsonNEU

Description

@WJohnsonNEU

image

Per programming manual for Keysight 34461A (emphasis mine):
The CONFigure commands select the function, range and resolution in one command. Specify
in the measurement's units (V, A, Hz, Ω, and so on). All other parameters are set to their
default values.

The defaults table is included in the picture above.

This should be included in the method's documentation so that using the method does not have unintended effects.